Hochschule
Karlsruhe Technik und Wirtschaft |
Requirement List Project
|
Summer Semester 2021 |
Faculty of Mechanical and Mechatronics |
|
|
Prof. J. Walter |
|
Semester:
SS 2021 |
Organisation- Details |
Process- Details |
|
Value-Details |
Num. |
Name: |
Art |
Phase |
Requirements
|
Minimum-
Fulfillment |
Should-
Fulfillment |
Ideal-
Fulfillment |
Unit of Measurement |
|
|
|
|
Physical-Technical Function
|
|
|
|
|
P01 |
Lorenz |
Y/N |
|
Measure load on each crutch |
+/- 10 |
+/- 5 |
+/- 2.5 |
%max partial load |
P02 |
Mufti |
W |
|
Measure crutch acceleration |
|
|
|
|
P03 |
Mufti |
W |
|
Acquire crutch orientation |
|
|
|
|
P04 |
Chris |
Y/N |
|
Recognize 3-point-walking with crutches |
|
|
|
|
P05 |
Chris |
W |
|
Recognize walking up- and downstairs with crutches |
|
|
|
|
P06 |
Chris |
Y/N |
|
Combine data between both crutches wirelessly |
|
|
|
|
P07 |
Mufti |
Y/N |
|
Calculate load on affected leg while walking |
+/-20 |
+/-10 |
+/-5 |
kg |
P08 |
Lorenz |
Y/N |
|
Support body weight |
70 |
|
|
kg per crutch |
P09 |
Joachim |
Y/N |
|
Provide continous battery runtime |
8 |
16 |
24 |
hours |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Technology (Manufacturability)
|
|
|
|
|
T01 |
Lorenz |
W |
|
Design to retrofit conventional crutch |
|
|
|
|
T02 |
Kai Li |
Y/N |
|
Meet ISO criteria |
|
|
|
|
T03 |
Lorenz |
W |
|
Design for future adaptation on walking frame (Zimmer frame) |
|
|
|
|
T04 |
Joachim |
Y/N |
|
Both crutches must be identical |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Economics
|
|
|
|
|
E01 |
Lorenz |
Y/N |
|
Maximale Gesamtkosten |
400 |
200 |
150 |
Euro |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Human-Product Relationships
|
|
|
|
|
H01 |
Joachim |
Y/N |
|
Signal patient overload acoustically |
|
|
|
|
H02 |
Joachim |
Y/N |
|
Signal patient overload optically |
|
|
|
|
H03 |
Danial |
Y/N |
|
Show live load on affected limb via
web application |
|
|
|
|
H04 |
Danial |
Y/N |
|
Imply basic operation for patient without additional devices |
|
|
|
|
H05 |
Kai Li |
Y/N |
|
Imply operation for medical personnel wireless via web application |
|
|
|
|
H06 |
Kai Li |
Y/N |
|
Set partial load (full load) via web application |
10-20 |
>0 |
|
kg |
H07 |
Mufti |
Y/N |
|
Store data anonymously on crutch |
|
|
|
|
H08 |
Joachim |
Y/N |
|
Simple recharge of battery |
|
USB Charger |
|
|
H09 |
Joachim |
W |
|
Handling for patient as simple as possible |
only calibration and on/off switches |
only on/off switch |
no switch |
|
H10 |
Danial |
W |
|
Show statistical data on web application: steps total, persentage, how many steps over/under limit, max/min load, average load, average above/below limit load |
|
|
|
|
H11 |
Danial |
W |
|
Show report in printable format |
html website |
PDF |
|
|
H12 |
Chris |
Y/N |
|
Determine patient's weight for calculation |
enter via web interface |
put whole weight on crutch, optional calibration switch |
calculate from raw data |
|
H13 |
Danial |
W |
|
Seperate display to show live the
load on affected limb without any operation |
|
|
|
|
H14 |
Kai Li |
W |
|
2nd operating mode that signals
load on affected limb below certain value |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Requirement Type : Y/N
- Necessary, TR - Tolerated Requirement; W - Wish; Construction Phase : P - Principle;
C - Concept; D - Draft; E - Elaboration;
|
First Edition
from April 2021 |
|
Edition: 18 April 2021 |
|
Version: 01 |
Editor:
|
|
|
|
Sheet 1 of 1 |